4-2. Operation of Actual Load Calibration

To carry out the actual load calibration, follow the procedures given below.
Load cell connection
Warm-up
Lock release
Host connection
Decimal place
Capacity
Minimum scale division
Balance weight
Zero calibration
Span calibration
Calibration lock
CAUTION
Once calibration is performed, the previous calibration values are lost and
cannot be restored. Be sure to turn ON the calibration lock except when
performing calibration.
Connect the F156 and load cell according to the section
"Connection."
Release the calibration lock which locks calibration.
Establish a CC-Link network.
(Connect the remote console unit if used.)
Register the position of Decimal point.
Register the maximum capacity of the scale.
If the registered value is exceeded by 10 scale divisions or more,
display shows over scale "OFL2".
Register the minimum scale division of the scale.
Register the value of load (balance weight) for calibration.
Calibration the initial zero point.
Place the load (balance weight) on the load cell and calibrate the
span value.
Disable changing set value related to calibration.
4.CALIBRATION
37