5.CALIBRATION

5-4-6. Minimum scale division

The minimum scale division can be from 001 to 100.
The display resolution obtains from dividing minimum scale division by capacity.
Operation
Press
MODE
The display resolution ignores the decimal place.
For example : When minimum scale division is 001 then
if capacity is 10.000, the display resolution is 1/10000;
if capacity is 100.00, the display resolution is 1/10000;
if capacity is 40.000, the display resolution is 1/40000.

5-4-7. Balance weight

The load value (Balance weight) that is applied to the scale for calibration. The input range of Balance
is between 0 to 99999.
Operation
Press
MODE
Input 0 ~ 99999

5-4-8. Gravitational acceleration

Slight error may occur if the scale moved from the location of calibration due to gravitational changes.
If the scale is used in the location of calibration, this function is not necessary. But after calibration the
scale will ship to different location, before Span calibration do operation procedure as follows.
- Select GRAV. ACCELERATION , input relative number to located Gravitational Acceleration.
- Do Span calibration.
- Input the number following new place Gravitational Acceleration.
Operation
Press
MODE
Input 01 ~ 16
01
9.806
07
9.800
13
9.794
32
CALIBRATION
MIN.SCALE DIVISION
CALIBRATION
BALANCE WEIGHT VALUE →
OK
CALIBRATION
PAGE
OK
Gravitational accceleration
02
9.805
03
9.804
08
9.799
09
9.798
14
9.793
15
9.792
→ Input 1 ~ 100
GRAV.ACCELERATION
04
9.803
05
10
9.797
11
16
9.791
OK
9.802
06
9.801
9.796
12
9.795

Advertisement

Table of Contents
loading

Table of Contents