6-3. Digital Zero Regulation Value

Set a range of zero point adjustment (a gap from the registered Zero Calibration value) by
digital zero or zero tracking. If digital zero operation is performed or zero tracking is actuated
where the digital zero regulation value is exceeded, the "zero alarm" status turns on, and the
digital zero regulation value is subtracted from the gross weight.
Setting range / 0000 to 9999
CC-Link
MSB
H
L
Console
1) Select code 45.
2) Press the ENT key, and then set the
zero regulation value.
3) Determine the setting with the ENT key.
Write the code No. and set value.
0
Code No.
00000 to 99999
Set value
Initial value / 200
Word size
remote register
LSB
RWw0006
45
RWw0005
RWw0004
Zero regulation value
(0000 ~ 9999)