Operating the device and sequence of reanimation

7.7 Opening SavePads and positioning electrodes

Fig. 11: Removing the film from the electrodes
(1) Protective film on electrodes
(2) SavePads adhesive electrodes
The PRIMEDIC™ HeartSave PAD will give spoken instructions for you to
apply the defibrillation electrodes to the patient. < Apply electrodes to patients bare
chest >
Procedure:
Open the defibrillation electrodes bag by tearing open the protective cover along
the tear strip.
Remove the protective film (1) from one of the electrodes (2) and then
immediately place the electrode on the position you had ascertained previously.
Refer to label stuck on the back of the electrodes.
Proceed to remove the protective film from the second electrode and place it in its
position.
Smooth the electrodes onto the patient ensuring there are no air bubbles under
the electrodes! Display < Apply electrodes to patients bare chest > must go out.
Operating instructions
HeartSave PAD 22289 / GB / C
DANGER
Do not touch the floor, other objects, clothing or other parts
of the body with the electrodes (after removing the
protective film) as it may remove the conducting layer of gel
on the electrodes.
A reduced layer of gel could cause burns on the skin under
the electrodes during defibrillation.
Please make sure that the red electrodes symbol LEDs on the
membrane keyboard go out.