Symbol
Description
rhythm
6
Patient with
Begin resuscitation (BLS)
Asystole
Not requiring defibrillation
rhythm
* CPR (Cardio Pulmonary Resuscitation) = HLW (Cardiopulmonary resuscitation)
Note
The scenario can be interrupted by pressing the  key on the keyboard. When
pressing the  key again, the scenario is continued at the same place.

4.2.2 Pediatric Mode

In barrel mode, there is the possibility to adjust the relationship between cardiac massage and
ventilation to 30:2 or 15:2.
Changing the ratio between cardiac massage and ventilation:
 Pressing the 9 button (7) changes from 15:2 / 30:2
Note
Ratio 30:2  2 x Beep
Ratio 15:2  1 x Beep
The ratio can be changed for Pediatric Mode!
The voice announcement 30 x chest compressions is then replaced by 15 x
chest compressions.
The setting is stored until the power module is removed, the next time the
trainer uses the last setting.
Note
During the analysis is not possible to change the mode
Note
The setting is saved until the energy module is removed.
METRAX GmbH
Resuscitation Sequence
2 min CPR
Analyse, No shock required
2 min CPR
Analyse, Shock required
2 min CPR
Analyse, Shock required
2 min CPR
Analyse, No shock required
Begin resuscitation
Analyse, No Shock required
23560 HeartSave AED Trainer / GB / B02
Using the device
17 / 24