2.7

Removing the microtome

2.8

Maintenance

Replacing the fuses
Replacing the fluorescent lamp
Leica CM1520 – Cryostat
• Do not use organic solvents or other aggressive substances for cleaning.
Use only disinfectants listed in these Instructions for Use, such as Leica
Cryofect (alcohol or other common alcohol-based disinfectants).
For additional detailed information about disinfection, contact
Leica Biosystems.
• The microtome is encapsulated and therefore does not require removal
by the user.
• Turn the instrument off and disconnect the power plug before replacing
the fuses.
• Use only the fuse types specified in Chapter 3, "Technical Data". Failure
to comply with this can cause severe damage to the instrument and its
surroundings!
• Turn the instrument off and disconnect the power plug before replacing
the lamps. (More details are in Chapter 9, "Cleaning, Disinfection, Main-
tenance.)