4.3.1 Removing or installing the transport anchors
28
1. Remove the transport anchor (24) for the Pick & Place module.
2. Remove the transport anchor (25) for the gripper. The gripper slowly moves downward in the
process.
• Using the provided No. 3 Allen key, unscrew the screws (22) and (23) for the two red transport
anchors (24) and (25)
24
22
23
25
• To transport the instrument, reinstall the transport anchors in the opposite order.
We recommend keeping the transport anchors and screws close to the instrument in the
provided clear plastic bag.
Leica CV5030 – Robotic Coverslipper
50
(see Fig.
8):
Transport anchor (24) for
the Pick & Place module
Transport anchor for the
gripper (25)
• Open the service door (28) on the left
side of the instrument and remove the
foam (50).
Fig. 7
23
4.
Instrument Setup
24
25
Fig. 8
21