● In addition, a high-frequency leakage current occurs as an inverter is used inside. When using a
ground-fault interrupter other than the specified ones with measures already taken against high
frequency, use one with a sensed current of 200 mA or more.
Table 8: List of power capacities and ratings of recommended ground-fault interrupters
Rating List
Recommended type
Rated current ( A )
Rated sensed current (m A )
Power capacity ( kVA )
■ Pin assignment of power
Figure 1:
Pin assignment
Table 9: Receptacle connector specifications
Manufacturer of receptacle
connector
Receptacle connector type
Applicable plug type
(Direct type)
Applicable electric wire size
Rated current (50/60 Hz)
After connecting the power connector, install the guard.
Screw
200V
NV-32SV(F)-( Manufactured by Mitsubishi Electric )
30
Pin assignment
Pin
No.
A
B
C
D
DDK Ltd.
CE05-2A22-22PD-D
CE05-6A22-22SD-D-BSS
Up to 10AWG/4 ( Outer diameter of insulation coating: φ4.5 mm )
46A
Power connector
Figure 14: Installation diagram of power connector guard
400V
30
9.5
Specifications
Phase L1/R
Phase L2/S
Phase L3/T
PE/GND
Japan Aviation Electronics Industry, Ltd.
JL04HV-2E22-22PE-B-R
JL04V-6A22-22SE-EB-R
Power connector guard
30
YO20-0938-DI-001-00
15