ITech IT8511A Manual page 65

Programmable electronic load
Table of Contents

Advertisement

th
26
byte
NOTE
Represent calibration protection state by one byte. Each byte is defined as:
From high to low
7 6 5 4 3 2 1 0
0 byte: protection state, 0 represent not in the protection state , 1 represent
in protection state.
38. Calibrate voltage value(62H)
st
1
byte
nd
2
byte
rd
3
byte
th
4
byte
th
th
From 5
to 25
byte
th
26
byte
NOTE
Current calibration standard points have four: 1, 2, 3,4.
39. Sending the actual input voltage to
calibration program(63H)
st
1
byte
nd
2
byte
rd
3
byte
th
4
byte
th
5
byte
th
6
byte
th
7
byte
th
th
From 8
to 25
byte
Sum code
Start bit ( AAH )
Address (0—0XFE)
Command(62H)
Voltage calibration point(1~4)
System reserve
Sum code
Start bit ( AAH )
Address (0—0XFE)
Command(63H)
The lowest byte of actual voltage
The lower byte of actual voltage
The higher byte of actual voltage.
The highest byte of actual voltage.
System reserve
65