4
LOCAL
LOCAL button state (0 is represent "not in effect ",1 is represent 'in effect ")
5
SENSE
Remote testing mode
6
LOT
FOR LOAD ON timer state
Demand state register
0
RV
Input reverse voltage
1
OV
Over voltage
2
OC
Over current
3
OP
Over power
4
OT
Over temperature
5
SV
Not connect remote terminal
6
CC
Constant current
7
CV
Constant voltage
8
CP
Constant power
9
CR
Constant resistance
36. Enter the calibration mode(60H)
st
1
byte
nd
2
byte
rd
3
byte
th
4
byte
th
5
byte
th
6
byte
th
th
From 7
to 25
byte
th
26
byte
NOTE
If Load is not in protection state, users could do the calibration operation.
37. Getting the calibration mode state(61H)
st
1
byte
nd
2
byte
rd
3
byte
th
4
byte
th
th
From 5
to 25
byte
Start bit ( AAH )
Address (0—0XFE)
Command(60H)
Calibration mode select(0:disable;1:enable)
Calibration password(0X85H)
Calibration password(0X11H or 0X12H)
System reserve
Sum code
Start bit ( AAH )
Address (0—0XFE)
Command(61H)
Calibration protection state
System
64