1-6. Power Supply and Air Supply Specifications

1-6. Power Supply and Air Supply Specifications
w
Prepare the interior wiring which is near the FVS-1000 and easy to connect with the connectors of the
FVS-1000.
Power supply voltage
+5 %
For UC
: AC 120 V
_10 %
: AC 220 - 240 V ± 10 %, 50/60 Hz ± 0.5 %
For CE
Taking the above specifications into consideration, execute the electric supply construction for breaker etc.
Air supply specifications
Pressure
: 0.25 to 0.7 Mpa (2.5 to 7.1 kgf/cm
Flux
: 25 Nl/min or more
Relative humidity : 80 % or less
: Water, oil and dust should not be included. (Max. 40 µm)
Others
To supply the outside air from the pipe arrangement for supplying the high-pressure air, check the
pressure, flux and quality of the air, then satisfy the above specifications using a regulator or filter, and
then connect. When there is no pipe arrangement, to supply the stable air, it is recommended to use the
oil free compressor equipped with a tank.
n
. If the air of which the pressure is more than the above specified value is supplied directly, there is a
danger of damaging the equipments.
. If the air which includes the water, oil or dust is supplied, there is a possibility of causing an internal
damage. Be sure to supply the clean air which passed through the air drier and the mist filter.
. As for the equipment to supply the outside air, refer to the operation manual of it. Be sure to apply the
periodic maintenance (draining the water from the tank etc.).