PR Alarm is from "1 + the lower limit of PR
Alarm" to 300.
 Press the Up Arrow or Down Arrow button to
increase or decrease values.
 Choose PR Lo Alarm in menu, press the
Function button enter setup. The PR Lo
Alarm box changes from real line to broken
line. The adjustable range for the lower limit
of PR Alarm is from 0 to "the upper limit of
PR Alarm – 1".
 Press the Function button, confirm the alarm
range setup.
 Hi Alarm is always higher than Lo Alarm by
at least 1 bpm.
 Press the On/Off button to exit the menu, and
return to measurement state.

3.3.12 Patient ID No. Setup

The oximeter can support 100 patient IDs, and 300-hour
data storage.
When entering menu, press the Function button to set
ID (valid range is from 1 to 100). The ID display box on
the interface will change from real line to broken line.
- 47 -