synchronously flash with PR value.

5.2.6 Lower than Lo Alarm Limit

If the measured SpO
or PR value is lower than the Lo Alarm
2
(lower alarm limit), the oximeter gives a high alarm.
Here we take SpO
for example:
2
If the measured SpO
value is lower than the set SpO
Lo Alarm,
2
2
the oximeter gives a low SpO
alarm. (Make sure Alarm System
2
in menu is ON.)
A ↓ icon displays near SpO
value, which indicates the
2
measured value is lower than that of SpO
Lo Alarm, it will
2
synchronously flash with SpO
value.
2

5.2.7 Alarm Silence

If Alarm System in menu is ON, pressing the alarm silence
button, the audio alarm will be off for the pause period set by the
user, but the visual alarm is still active.
When the audio alarm is off, press the Alarm Silence button to
reactivate the audio alarm function.

5.2.8 Turning off Alarm System

After the alarm system is turned off, the oximeter can not give a
visual or an audio alarm except for low battery icon alarm.
Set Alarm System to ON, the alarm system will be active. It will
give an audio alarm and a visual alarm if an alarm occurs.
- 61 -