VE-H100B Veterinary Pulse Oximeter User Manual
Checking the Ni-MH Battery Package
The performance of a Ni-MH rechargeable battery package may
deteriorate. To check the performance of the battery, follow the
procedures below:
1.
Disconnect the pulse oximeter from the patient and stop all
monitoring and measuring procedures.
2.
Place the pulse oximeter in the charger stand and connect the
AC mains. Allow the battery to be charged uninterruptedly for
above 2.5 hours. For details about charging the Ni-MH battery
package, please refer to section 3.4.
3.
Disconnect AC mains and allow the pulse oximeter to run in
the measurement state until it shuts off.
4.
The operating time of a battery reflects its performance
- 30 -