Page 1
Operating Manual APT.line™ B 28 Incubators with mechanical control APT.line™ E 28 Heating ovens with mechanical control BINDER GmbH Address Post office box 102 D-78502 Tuttlingen Tel. +49 7462 2005 0 +49 7462 2005 100 Internet http://www.binder-world.com E-mail info@binder-world.com Service Hotline...
Page 2
EG - KONFORMITÄTSERKLÄRUNG EC - DECLARATION OF CONFORMITY CE - DECLARATION DE CONFORMITE Anbieter / Supplier / Fournisseur: BINDER GmbH Im Mittleren Ösch 5, D-78532 Tuttlingen Anschrift / Address / Adresse: Produkt / Product / Produit: Brutschränke mit mechanischer Regelung Incubators with mechanical control Incubateurs à...
Page 3
à la fonction des matériels portatifs d’essai, de mesure et de surveillance utilisés dans des systèmes de distribution basse tension. D-78532 Tuttlingen, 24.01.2011 BINDER GmbH P. M. Binder Dr. H. von Both Geschäftsführender Gesellschafter Leiter F & E Managing Director Director R &...
Page 4
EG - KONFORMITÄTSERKLÄRUNG EC - DECLARATION OF CONFORMITY CE - DECLARATION DE CONFORMITE Anbieter / Supplier / Fournisseur: BINDER GmbH Im Mittleren Ösch 5, D-78532 Tuttlingen Anschrift / Address / Adresse: Produkt / Product / Produit: Wärmeschränke mit mechanischer Regelung Heating ovens with mechanical control Étuves à...
Page 5
à la fonction des matériels portatifs d’essai, de mesure et de surveillance utilisés dans des systèmes de distribution basse tension. D-78532 Tuttlingen, 24.01.2011 BINDER GmbH P. M. Binder Dr. H. von Both Geschäftsführender Gesellschafter Leiter F & E Managing Director Director R &...
TEMPERATURE SAFETY DEVICE CLASS 1 (OPTION) ........18 MAINTENANCE, CLEANING, AND SERVICE ............ 19 Maintenance intervals, service......................19 Cleaning and decontamination ......................19 Sending the unit back to BINDER GmbH ..................21 DISPOSAL......................21 Disposal of the transport packing .....................21 Decommissioning..........................21 Disposal of the unit in the Federal Republic of Germany ..............22 Disposal of the unit in the member states of the EC except for the Federal Republic of Germany .23...
Understanding and observing the instructions in this operating manual are prerequisites for hazard-free use and safety during operation and maintenance. In no event shall BINDER be held liable for any damages, direct or incidental arising out of or related to the use of this manual.
WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious (irreversible) injury CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in moderate or minor (reversible) injury CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in damage to the product and/or its functions or of a property in its proximity.
Pictograms (Warning signs) Service label Hot surface Figure 1: Position of labels on the unit Keep safety labels complete and legible. Replace safety labels that are no longer legible. Contact BINDER Service for these replacements. B / E 01/2011 page 9/30...
1/119 laboratory guidelines issued by the employers’ liability insurance association) (for Germany). BINDER GmbH is only responsible for the safety features of the unit provided skilled electricians or qualified personnel authorized by BINDER perform all maintenance and repair, and if components relating to chamber safety are replaced in the event of failure with original spare parts.
Page 11
To operate the unit, use only original BINDER accessories or accessories from third-party suppliers authorized by BINDER. The user is responsible for any risk caused by using unauthorized accessories. CAUTION Danger of overheating. Damage to the unit. ∅ Do NOT install the unit in unventilated recesses.
BINDER incubators B and heating ovens E are heated electrically and are ventilated naturally. BINDER incubators B and heating ovens E are NOT regularly equipped with any temperature safety device consequently correspond to class 0 according to DIN 12880. Operating is allowed under supervision only.
CAUTION Danger of overheating. Damage to the unit. Damage to the charging material. ∅ Do NOT operate an unsupervised unit. Operate the unit under supervision only. The units can be optionally equipped with a temperature safety device class 1.0 according to DIN 12880 (chap.
E 28 unit overview Figure 4: E 28 heating oven Control thermometer Timer 0-120 minutes (also main power switch) Pilot lamp green (ready to operate) Thermostat Pilot lamp yellow (heating active) Thermostat brake Completeness delivery, transportation, storage, installation Unpacking, and checking equipment and completeness of delivery After unpacking, please check the unit and its optional accessories, if any, based on the delivery receipt for completeness and for transportation damage.
Second-hand units are units that were used for a short time for tests or exhibitions. They are thoroughly tested before resale. BINDER ensures that the chamber is technically sound and will work flawlessly. Second-hand units are marked with a sticker on the unit door. Please remove the sticker before commissioning the unit.
• Permissible ambient humidity: 70 % r.H. max., non-condensing. • Installation height: max. 2000 m / 6.6 ft. above sea level. When placing several units of the same size side by side, maintain a minimum distance of 250 mm / 9.84 in between each unit.
Start up Turning on the unit Warming chambers may release odors in the first few days after commissioning. This is not a quality defect. To reduce odors quickly we recommend heating up the chamber to its nominal temperature for one day and in a well ventilated location. 5.1.1 Turning on the B 28 incubator by the main power switch Figure 5: Main power switch ON/OFF (B 28) Set the main power switch (2) to position I.
If the heating elements turn off repeatedly, have a specialist investigate and remove the reason of the failure. The reset button is located on the unit rear, at the bottom right side. If the temperature safety device acts repeatedly, please contact an authorized service or BINDER Service. B / E 01/2011 page 18/30...
Ensure regular maintenance work is performed at least once a year. Replace the door gasket only when cold. Otherwise, the door gasket may become damaged. We recommend taking out a maintenance agreement. Please consult BINDER Service. BINDER telephone hotline: +49 (0) 7462 2005 555...
Page 20
We recommend using the neutral cleaning agent Art. No. Art. Nr. 1002-0016 for a thorough and mild cleaning. Any corrosive damage that may arise following use of other cleaning agents is excluded from liability by BINDER GmbH. B / E 01/2011 page 20/30...
Sending the unit back to BINDER GmbH If you return a BINDER product to us for repair or any other reason, we will only accept the product upon presentation of an authorization number that has previously been issued to you. An authorization number will be issued after receiving your complaint either in writing or by telephone prior to your sending the BINDER product back to us.
(Elektro- und Elektronikgerätegesetz, ElektroG) from 23 March 2005, BGBl. I p. 762 or contact BINDER service who will organize taking back and disposal of the unit according to the German national law for electrical and electronic equipment (Elektro- und Elektronikgerätegesetz, ElektroG) from 23 March 2005,...
According to directive 2002/96/EC of the European Parliament and of the Council on waste electrical and electronic equipment (WEEE), BINDER devices are classified as “monitoring and control instruments” (category 9) only intended for professional use“. They must not be disposed of at public collecting points.
(certified since December 1996 by TÜV CERT). All test equipment used is subject to the administration of measurement and test equipment that is also constituent part of the BINDER QM DIN EN ISO 9001 systems. They are controlled and calibrated to a DKD-Standard at regular intervals.
°F and a power supply voltage fluctuation of ±10. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width and depth of the inner chamber.
°F and a power supply voltage fluctuation of ±10. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width and depth of the inner chamber.
-- Not available Accessories and spare parts BINDER GmbH is only responsible for the safety features of the unit only, provided skilled electricians or qualified personnel authorized by BINDER perform all maintenance and repair, and if components relating to chamber safety are replaced in the event of failure with original spare parts.
Contamination clearance certificate Unbedenklichkeitsbescheinigung Declaration regarding safety and health Erklärung zur Sicherheit and gesundheitlichen Unbedenklichkeit The German Ordinance on Hazardous Substances (GefStofV), and the regulations regarding safety at the workplace, require that this form be filled out for all products that are returned to us, so that the safety and the health of our employees can be guaranteed.
Page 29
Kind of transport / transporter / Transportweg/Spediteur: Transport by (means and name of transport company, etc.) Versendung durch (Name Spediteur o.ä.) ___________________________________________________________________________________ Date of dispatch to BINDER GmbH / Tag der Absendung an BINDER GmbH ___________________________________________________________________________________ B / E 01/2011 page 29/30...
Page 30
We are aware that, in accordance with Article 823 of the German Civil Code (BGB), we are directly liable with regard to third parties, in this instance especially the employees of BINDER GmbH, who have been entrusted with the handling / repair of the unit / component. / Es ist uns bekannt, dass wir gegenüber Dritten –...