12.8. Error codes

When an error occurs, one of the following error codes is displayed in the systolic display
section.
Printer error codes
Error code
No printer paper. Install a new roll of printer paper.
The printer cover is open. Firmly close the printer cover.
A printer cutter error. Open the printer cover, check the printer paper, and then close
the printer cover.
Error code details
Error code
Error related to blood pressure measurement
When the power is switched
on, the pressure detection is
unstable.
An electrical error is detected
in the blood pressure
measurement section.
The safety monitor of the blood
pressure measurement section
detected an error.
,
Pressure is not applied at the
start of the measurement.
Pressure cannot be applied
within a certain period of time.
Inflation speed is too fast.
The exhaust speed is too slow.
The exhaust speed is too fast.
Error/countermeasure
Details
Check if there is air remaining in the cuff.
Restart and then try blood pressure measurement
again. If the problem continues, stop using the
monitor immediately.
Restart and then try blood pressure measurement
again. If the problem continues, stop using the
monitor immediately.
A condition that may affect the safety of the patient
was detected during measurement. External
vibrations may have been applied to the
air system of the cuff or inside the monitor or an
obstruction may have been mistakenly detected.
Check the patient condition and measurement
environment and try blood pressure measurement
again. If the problem continues, stop using the
monitor immediately.
There may be an air leak in the air system inside
the monitor. If the problem continues, stop using
the monitor immediately.
There may be a leak in the air system inside the
monitor or the cuff was applied loosely. If the
problem continues, stop using the monitor.
There may be a bend or blockage in the air
system inside the monitor. If the problem
continues, stop using the monitor.
Air is not being correctly exhausted. There may be
a bend or blockage in the air system inside the
monitor. If the problem continues, stop using the
monitor.
The patient may have moved or a strong external
pressure was applied during measurement. If the
problem continues, stop using the monitor.
Check items