Error code
Excess pressure was detected.
The time limit for one
measurement was exceeded.
The pressure is insufficient.
Pulse cannot be detected.
Diastolic blood pressure cannot
be determined.
Mean arterial blood pressure
cannot be determined.
Systolic blood pressure cannot
be determined.
Pulse cannot be determined.
The blood pressure value is
inappropriate.
SYS value is 'out of range'.
DIA value is 'out of range'.
PUL value is 'out of range'.
Details
The cuff pressure during measurement exceeded
300 mmHg.
The patient may have moved or a strong external
pressure was applied to the cuff.
Watch for errors and try measurement again.
For the safety of the patient, measurement was
cancelled because the measurement time
exceeded 180 seconds.
Measurement may have been repeated. Check
the patient for body movement and arrhythmia.
Blood pressure measurement was not possible
because the pressure was insufficient. During
inflation, patient movement or an external vibration
introduced noise into the cuff pulse and the set
pressure was detected or the patient's blood
pressure rose greatly during blood pressure
measurement.
Confirm the following conditions:
The cuff is not loose; no thick clothing on the arm; the
patient remains still; and no external vibrations on the
cuff. And try measurement again.
The pulse signal received by the cuff is too low.
The circulation of the patient may be poor or the
patient is wearing thick clothing.
Check the condition of the patient.
Check the patient for body movement and
arrhythmia.
SYS measurement range : 40-270 mmHg
Check the patient for body movement and
arrhythmia.
DIA measurement range : 20-200 mmHg
Check the patient for body movement and
arrhythmia.
PUL measurement range : 30-240 mmHg
Check the patient for body movement and
arrhythmia.
Check items
53