7. Be aware that strong electromagnetic waves can cause malfunctions.
This monitor complies with EMD-standard IEC60601-1-2:2014. However, to prevent
electromagnetic interference with other devices, do not use mobile phones close to
within 30 cm of near the monitor.
If this monitor is located near strong electromagnetic waves, noise may enter in
waveforms and malfunctions may occur. If unexpected malfunctions occur during
use of this monitor, inspect the electromagnetic environment and take appropriate
actions.
The following are examples of general causes and countermeasures.
Use of mobile phones
Radio waves may cause unexpected malfunctions.
Instruct visitors to rooms or buildings with medical electrical devices not to use mobile
phones or small wireless devices.
High frequency noise is being introduced from other devices via the electrical outlet.
Check for the source of noise, and then perform countermeasures, such as using a
noise cancellation device on this line.
If the noise source is a device that can be stopped, stop using it.
Use another electrical outlet.
Effects from static electricity are suspected (discharges from devices or the
surrounding area)
Before using the monitor, ensure that the operator and patient have discharged static
electricity.
Humidify the room.
If lightning is occurring nearby, the monitor may receive excessive voltage. In such
cases, power the monitor using the following method.
Use an uninterruptible power supply.
8. Environmental considerations
Before disposing of this monitor, remove the lithium battery from the monitor.