Electromagnetic compatibility
Immunity tests
Voltage dips,
brief supply volt-
age interruptions
and fluctuations
according to
IEC 61000-4-11
Magnetic field at
supply frequency
(50/60 Hz)
according to
IEC 61000-4-8
Conducted RF
interference
according to
IEC 61000-4-6
1)
UT is the AC mains voltage prior to test level application
50
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 50
Test level
EN 60601-1-2
EN 60601-2-24
< 5% UT ¹
for ½ periods
(>95% dip)
40% UT ¹
for 5 periods
(60% decline)
70 % UT ¹
for 25 periods
(30 % decline)
<5% UT ¹
for 5 s
(>95% dip)
3 A/m
IEC 60601-1-2:
150 kHz to 80 MHz
3 V
outside and
RMS
10 V
in ISM
RMS
frequency bands
IEC 60601-2-24:
150 kHz to 80 MHz
10 V
RMS
Compliance
Electromagnetic
level
environment guidelines
Complies
The supply voltage quality
through
should be the same as that
the use of
of a typical commercial or
an internal
hospital environment.
energy source
400 A/m
Magnetic fields at the
supply frequency should
correspond to those typi-
cally found in commercial
and hospital environments.
10 V
Do not use portable radio
RMS
150 kHz to
communications equipment
80 Mhz
closer to the Infusomat
in all
compact
frequency
connection cables) than
bands
the recommended safe
distance calculated using
the appropriate equation for
that frequency.
Recommended safe
distance: d = 1.2 √P
plus
P (including
1)
15.10.2020 17:47:32