Technical data
Parameter
KVO rate
Air detector
History protocol
Note:
The preset bolus rate (800 ml/h) can be changed via the service menu or once via the
combination of bolus volume and bolus time. Delivery accuracy during bolus administration
is typically ±5%. The accuracy can deviate when administering small bolus volumes.
Essential performance characteristics for infusion pumps
Infusion of liquids without variation of infusion rate
Q
Pressure limitation as protection from the bursting of the infusion line
Q
Protection from air-infusion
Q
Protection against unintended bolus volumes and occlusion (added by IEC 60601-2-24)
Q
Alarm signal of high priority (added by IEC 60601-2-24)
Q
46
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 46
Value
Rate: ≥ 10 ml/h: KVO rate 3 ml/h
Q
Rate: < 10 ml/h: KVO rate 1 ml/h
Q
Rate: < 1 ml/h: KVO rate = rate set using the service
Q
program (factory default rate 0.1 ml/h) or current rate if
this is lower.
Technical sensitivity:
Detection of air bubbles ≥ 0.01 ml.
Alarm trigger:
Individual air bubble alarm: 0.02 – 0.3 ml
(standard 0.3 ml)
Cumulative air alarm: 0.5 – 3.8 ml/h
(standard 1.5 ml/h)
Trigger: 0.01 ml
1,000 history entries
Q
The oldest entries are overwritten if necessary.
100 events for system diagnosis
Q
The history is retained when the device is switched off or
the battery removed.