Start-up and trumpet curves
These graphs show the accuracy and
uniformity of flow over time. Take into
account:
The delivery behaviour and the delivery
Q
accuracy are fundamentally affected
by the disposable item used.
Variations in consumable and consum-
Q
able handling (e.g. stretching) may
affect accuracy.
A container placed below the pump
Q
may result in delivery deviations.
The system accuracy is normally ± 5 %
Note:
of the volume, measured using the trumpet
curve test method according to IEC 60601-
2-24 at a rate of 25 ml/h (at 20 °C ± 2 °C)
and using the recommended disposable
item.
Trumpet curves
(Measured values for second hour in
each case)
Measurement
interval
Observation
interval
Start-up curves
Measurement
interval
Measurement
duration
Flow Q
i
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 43
Δt = 0.5 min
p x Δt [min]
Δt = 0.5 min
T = 120 min
(ml/h)

Alarm times

16.3
The following graphics show alarm times
depending on pressure level.
Note:
length of disposable and temperature
might affect the time to alarm.
Intrafix® Primeline
16.3.1
Zeit
Time [h:mm:ss]
(hh:mm:ss, logarithmisch)
24:00:00
0,1 ml/h
0.1 ml/h
02:25:00
02:20:26
1 ml/h
1 ml/h
00:25:00
00:08:25
00:02:00
25 ml/h
25 ml/h
00:00:10
00:00:07
00:00:01
P1 =
P1 =
Pressure level
Druckstufe
67 mBar
67 mBar
21:02:35
01:14:09
00:01:50
Intrafix® Primeline
P9 =
P9 =
1000 mBar
1000 mBar
43
15.10.2020 17:47:31