Set-up and start-up
Turning night mode on/off
7.3.1
In night mode the display brightness is
reduced.
Select Night mode and press OK to
Q
confirm.
Select On / Off and press OK to confirm.
Q
Setting display brightness
7.3.2
Select Brightness and press OK to
Q
confirm.
Select brightness level and press OK
Q
to confirm.
Level 1 (=lowest level)
–
- to -
Level 9 (=highest level)
–
Setting the audio volume
7.3.3
Select Audio volume and press OK to
Q
confirm.
Select Audio volume level and press
Q
OK to confirm.
Level 1 (=lowest level)
–
- to -
Level 9 (=highest level)
–
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 23
Configuring the pressure
7.3.4
alarm limit
WARNING! Danger to the patient
from an incorrectly set pressure
alarm limit.
Ensure that the pressure alarm level
Q
limit is set so that the alarm can be
triggered in good time.
It may be necessary to change the pressure
alarm limit due to various influencing fac-
tors, e.g. temperature, line length and inner
diameter and the filter used in the system
set-up.
Note:
The set pressure level affects the time
to alarm. In order to minimize the time to
alarm, it is recommended that you start
with a low pressure level and increase it if
required. The set pressure level affects the
alarm time.
Note:
In the event of a pressure alarm, the
post occlusion bolus will be automatically
reduced.
Select Pressure alarm and press OK to
Q
confirm.
Select alarm level and press OK to
Q
confirm.
Level 1 (=lowest level)
–
- to -
Level 9 (=highest level)
–
Alarm
Pressure value
level
1
0.067 bar (50 mmHg)
2
0.133 bar (100 mmHg)
3
0.200 bar (150 mmHg)
23
15.10.2020 17:47:27