Electromagnetic compatibility
Electromagnetic immunity
18.2
The device is designed to be used in the electromagnetic environmental conditions described
below. Customers or users of the compact
the system is being operated in such an environment.
Immunity tests
Electrostatic
discharge (ESD)
according to
IEC 61000-4-2
Electrical fast
transient/ bursts
according to
IEC 61000-4-4
Surges according
to IEC 61000-4-5
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 49
plus
Test level
EN 60601-1-2
EN 60601-2-24
Contact discharge
IEC 60601-1-2:
±6 kV
IEC 60601-2-24:
±8 kV
Air discharge
IEC 60601-1-2:
±8 kV
IEC 60601-2-24:
±15 kV
for power supply
lines ±2 kV
For input and output
lines ±1 kV
Differential mode
voltage ±1 kV
Common mode
voltage ±2 kV
system or its components should ensure that
Compliance
Electromagnetic
level
environment guidelines
Floors should be wood,
±6 KV
concrete or ceramic tile.
without
If floors are covered with
interference
synthetic material, the rela-
±8 KV outage
tive humidity should be at
with alarm
least 30 %.
permitted
±8KV without
interference
±15KV outage
with alarm
permitted
±2 kV
Mains power quality should
be that of a typical com-
mercial or hospital environ-
±1 kV
ment.
±1 kV
Mains power quality should
be that of a typical com-
mercial or hospital environ-
±2 kV
ment.
49