Technical data
Parameter
Acoustic alarm signal
sound pressure range
Interfaces
Operating conditions
Temperature
Q
Relative air humidity
Q
Atm. pressure
Q
Storage conditions
Temperature
Q
Relative air humidity
Q
Atmospheric pressure
Q
Weight
Dimensions in mm
(W x H x D)
Safety check
Volume preselection
Time preselection
Delivery accuracy
Occlusion alarm pressure 9 levels up to 1 bar
Alarm in the case of
incorrect dose
Delivery rate increments
Delivery accuracy for
bolus administration
Max. bolus volume after
bolus reduction
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 45
Value
Nine available levels:
45 dB(A) to 75 dB(A)
Cold connector for mains voltage
Q
Accessory port for interface cable 12 V CP and staff call
Q
IrDA infrared for communication in the station and for
Q
service
+18 °C ... +35 °C / +64 °F ... +95 °F
Q
30% ... 90% (without condensation)
Q
0.54 ... 1.06 bar
–20 °C ... +55 °C / -4 °F ... +131 °F
Q
20% ... 90% (without condensation)
Q
0.5 ... 1.06 bar
Q
Approx. 2.0 kg
Approx. 229 mm x 98 mm x 225 mm
(including pole clamp compact
Every 2 years
0.1 ml - 9,999 ml in increments of 0.01 ml
00:01 h - 99:59 h
typ. ± 5% according to IEC 60601-2-24
Note:
valid for 50 cm water column
In the event of an incorrect dose of 1 ml due to pump
malfunction (electronics, software), the pump will automati-
cally switch off.
0.1 ml/h ... 1000 ml/h in increments of 0.01 ml/h
typ. ± 5% for bolus volumes > 5 ml
≤ 0.2 ml at rates ≥ 1ml/h
plus
)
45