B. Braun Infusomat compact plus P Instructions For Use Manual page 51

Valid for software 003a
Table of Contents

Advertisement

Electromagnetic compatibility
Immunity tests
Radiated RF
interference
according to
IEC 61000-4-3
2)
With P as the maximum rated power of the transmitter in watts (W) according to the trans-
mitter manufacturer specifications and d as the recommended safe distance in metres (m).
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 51
Test level
EN 60601-1-2
EN 60601-2-24
10 V/m
80 MHz to 2.5 GHz
Compliance
Electromagnetic
level
environment guidelines
10 V/m
The field strength should be
80 MHz to
lower than 10 V/m
6 GHz
d = 1.2 x √P
80 MHz to 800 MHz
d = 2.3 x √P
800 MHz to 6 GHz
Field strengths from
stationary RF transmit-
ters should be below the
compliance level for all
frequencies, based on an
on-site test.
Interference is possible in
the vicinity of equipment
that has the following
symbol.
2)
2)
51
15.10.2020 17:47:32