Charging The Battery; Replacing Sensors - TESTO 320 Instruction Manual

Flue gas analyzer
Hide thumbs Also See for 320:
Table of Contents

Advertisement

7 Maintaining the product
7.3.
7.4.
60
info@Testo-Direct.com
All manuals and user guides at all-guides.com

Charging the battery

The rechargeable battery can only be charged at an ambient
temperature of ±0 to +35°C. If the rechargeable battery has been
completely discharged, the charging time at room temperature with
the testo mains unit is approx. 6 h.
Charging in the measuring instrument
1. Connect the mains unit instrument plug to the instrument's
micro USB socket.
2. Connect the mains plug of the mains unit to a mains socket.
-
The charging process will start. The charging process will stop
automatically when the battery is fully charged.
Battery care
> Do not fully exhaust rechargeable batteries.
> Store rechargeable batteries only in charged condition and at
low temperatures, but not below 0°C (best storage conditions
with a charge level of 50-80%, at an ambient temperature of
10 - 20°C, recharge completely before use).

Replacing sensors

✓ The measuring instrument must be switched off.
1. Place the measuring instrument on its front.
2. Unscrew, lift up and remove the service cover.
3. Disconnect the hose connections from the faulty sensor/bridge.
4. Remove the faulty sensor/bridge from the slot.
Testo-Direct
www.
1.888.610.7664
.com

Advertisement

Table of Contents
loading

Table of Contents