Technical data
3
Technical data
Description
Instrument height
Instrument width
Instrument depth
Instrument weight
Power supply
Power during measurement
Power during stand-by
Supply voltage fuse
Operating range (permissible ambient
temperature)
Recommended temperature range to
achieve the specified performance
Place of installation
Air humidity range
Sound pressure level
— 20
Value
Benchtop instrument: 450 mm; 17.7"
Floor instrument: 1350 mm; 53.1"
Benchtop instrument: 625 mm; 24.6"
Floor instrument: 625 mm; 24.6"
Benchtop instrument: 750 mm; 29.5"
Floor instrument: 790 mm; 31.1"
Benchtop instrument: approx. 78 kg; ~ 172 lbs
Floor instrument: approx. 132 kg; ~ 291 lbs
90 - 253 VAC, 50/60 Hz
400 VA
180 VA
5 A (slow-blow)
10°C – 30°C; 50°F – 86°F
18°C – 28 °C; 64°F – 82°F
Inside of dry, dust-free rooms up to an altitude of
2000 m above sea level, vibration-free (not
subject to impact or vibration load conditions)
20 – 80%, not condensing
55 dB(A)