5
Initial start-up
5.1
Preparing the instrument for use
Before switching on the instrument, proceed as follows:
1.
Check whether the power supply of the instrument supplies the correct input
voltage (90 - 253 VAC, 50/60 Hz).
2.
Check if the voltage selector switch of the PC is set to the proper input
voltage (230V/115V).
3.
Check the filling level of the gas-washing bottle. Fill with water if required, see
page 47.
4.
Set the argon pressure to 4.8 - 5.0 bar (69.6 - 72.5 psi).
If the argon pressure is not properly set, turbulences result in the spark
stand insert. Thus, correct measuring results cannot be obtained.
The instrument is ready for use.
SPECTROMAXx — 21.02.2020
Initial start-up
— 25