Troubleshooting
Malfunction
The instrument cannot be
switched on.
The burn spot is
incorrect.
SPECTROMAXx — 21.02.2020
Cause
The power plug is not
plugged in properly.
Power is not available.
Defective fuse.
The argon outlet (exhaust)
is clogged with
condensate.
Insufficient argon supply.
Poor argon quality.
Incorrect electrode gap.
Plasma generator faulty.
The samples have not
been prepared properly.
The sample surface is not
flat enough.
Troubleshooting
Remedy
Check if the power plug was
properly plugged in.
Check the power supply.
Check the fuse inside the power
connection socket. If necessary,
replace the fuse.
Clean the spark stand including
the argon outlet.
Check argon supply and
pressure.
Connect new argon;
specifications, chapter 3
„Technical data" on page 20.
Clean the electrode and re-adjust
the gap using the spacer.
Please contact your Service
Center.
Prepare sample properly, (grind,
turn or mill).
— 53