Troubleshooting
Malfunction
Low intensities in UV
optics.
Low intensities in the air
optics.
Limit values are not
maintained during
iCALising.
— 54
Cause
Light inlet window soiled.
Light entrance lens is
soiled.
Shutter blocked.
Low transmission.
Light inlet window soiled.
Insufficient intensity.
Intensities too high.
Remedy
Clean light inlet window if
required, see page 37.
Please contact your Service
Center.
Check shutter function, call the
Service Center if required
Use argon to rinse the optics
(function "Rinse" or the
corresponding function key)
Clean light inlet window if
required, see page 37.
1. Check argon.
2. Check sample.
3. Clean spark stand adjust
electrode.
4. Clean light inlet window.
5. ICALize the instrument again
and, if necessary, contact the
Service Center.
1. Check the shutter function.
2. Check the plasma generator
parameters, contact the
Service Center if necessary.
SPECTROMAXx — 21.02.2020