Replacing the ICALization sample

When the ICALizing sample is almost used up, a replacement must be provided.
The composition of the ICALizing sample is different from that of the original
samples. The intensities of the new samples must be matched to those of the old
samples.
Please contact your Service Center.
Information
Do not discard the old ICALization sample before the intensities have been
matched. Matching is not possible without using the old samples.