Making A Ovemp Recording; Electrode Montage For Ovemp; Stimuli For Ovemp; Default Collection Parameters - Interacoustics Eclipse EP15 Instructions For Use Manual

Table of Contents

Advertisement

4.7 Making a oVEMP recording

The oVEMP test is a test to assist in the assessment of vestibular function.

4.7.1 Electrode montage for oVEMP

Red
Under the left eye (reference)
White
Chin (active)
Black
Ground, forehead (common)
Blue
Under the right eye (reference)

4.7.2 Stimuli for oVEMP

Typically, a 500Hz tone burst is used for oVEMP testing as it provides the largest response amplitude. Other
stimuli available for testing include:
Tone bursts 250 Hz – 4 kHz
Click
Custom WAV file (when Research Module license is enabled)

4.7.3 Default collection parameters

The standard oVEMP protocol, settings for the stimulus and recordings are:
Stimulus parameters
Type: 500 Hz ToneBurst, Blackman (2:2:2)
Rate: 5.1/s
Intensity: AC: Typically ≥ 95 dBnHL is used, but is not preselected, and must be set by the user prior to
staring the test.
Recording parameters
Low pass filter: 1000 Hz
High pass filter: 10 Hz 6/oct
Electrode impedance less than 5 kohms, with interelectrode impedance less than 3 kOhm.
Recording epoch 100ms with a 10ms prestimulus recording period and a 90ms post stimulus recording
period.
Approximately 150 sweeps per waveform shall be collected.
D-0120588-B – 2019/03
Eclipse – Instructions for Use - EN
Page 65

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Eclipse EP15 and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Eclipse ep25Eclipse vempEclipse assrEclipse abris

Table of Contents

Save PDF