Datcon DT9000 Operating Instructions Manual page 69

Intrinsically safe process indicator
Table of Contents

Advertisement

DT9000

Access is possible
by supervisor code only

Access is possible
by supervisor code only

Access is possible
by supervisor code only

Access is possible
by supervisor code only

20171205-V2
Physical value assigned to 4 mA
The physical value to be assigned to the 4 mA loop current.
Any values can be defined for it within the minimum-
maximum interval that can be displayed . [Default factory
setting: 4 000]
Physical value assigned to 20 mA
The physical value to be assigned to the 20 mA loop
current. Any values can be defined for it within the
minimum-maximum interval that can be displayed. [Default
factory setting: 20 000]
Averaging number (Averaging Number)
The displayed measurement result is generated as the
average of several measurements. The number of
measurements can be defined here.
By increasing this number the display stability increases
too, but the signal settling time becomes lower.
[Default factory setting: 8]
Display refresh time (Display Refresh)
The instrument performs cca. 15-16 measurements in each
seconds. It's too fast to see the changing of the
measurement value.
Here you can define the time periods by which the
instrument displays the new measurement values.
[Default factory setting: 0.5 sec]
Test programmes sub-menu (Test)

Display testing (Test: LCD)
Checking if the display works correctly.

Output testing (Test: Output)
Displaying and changing the status of the limit outputs,
independently of their operating modes.
69