10 PC Diagnostic Tool
10.4 Troubleshooting
Test
Processor(s)
Motherboard testing
Stress test drive
Memory testing
2D test
Multimedia testing
USB testing
86
10.4.2

Sirona Windows Diagnostics

Test Description
Continuous load and inspection of pro-
cessor cores
Continuous load and inspection of the
motherboard functions
Continuous load and inspection of the
hard disk
Continuous load and inspection of the
memory
Continuous load and inspection of 2D
properties of the graphics card
Continuous load and inspection of 3D
properties of the graphics card
USB test
(at least one USB human device must be
connected, such as a keyboard, trackball,
camera, or UPS mains supply)
Operating Instructions CEREC AC Connect
Action if problems occur
➢ Replace the PC.
➢ Replace the PC.
1. Check the hard disk's SATA line.
2. Check the hard disk's power supply.
3. Exchange the hard drive including
the hard drive cable.
➢ Remove the memory and then re-
insert it.
➢ Replace the graphics card.
➢ Replace the graphics card.
➢ Replace the PC.