4 Operation
4.5 Accessories for the operating light
36
4.5.2

Patient mirror

General information on the patient mirror
The accessory interface must be installed on the operating light in order
to use the patient mirror.
You can install the patient mirror on the mounting adapter of the
accessory interface and remove it again when not in use.
CAUTION
The operating light can be accidentally switched off via the contactless
sensor and the patient dazzled when using the mirror in the proximity of
the mouth.
Before using the mirror in the proximity of the mouth, switch the
operating light off and switch off the contactless sensor function via the
treatment center keyboard or via the integrated keyboard (see sections
"Operation via dentist or assistant element" [ → 22] or "Operation via
integrated keyboard" [ → 31]).
Mounting the patient mirror
The accessory interface is installed on the operating light.
1. Set the bayonet mount of the patient mirror at a 45° angle at the
mounting adapter.
2. Turn the patient mirror clockwise until it locks into place.
3. Check that the patient mirror is securely positioned.
Dentsply Sirona
Operating Instructions LEDview Plus
65 67 510 D3642
D3642.201.01.03.02 01.2018