Replacing the color calibration set
Please note that the color calibration set
● can only be used with CEREC SW software ≥ 4.5 or Sirona Connect
SW ≥ 4.5 or CEREC Premium SW ≥ 4.5.
● can only be kept for use for a maximum of 2 years. You can find the
expiry date at the bottom of the color calibration set container.
Previous storage may mean that the period for use has been
reduced to less than 2 years.
● can only be used for one year after the container has been opened.
Write the date that the container was opened on the container after
"Opened on ___________" using a waterproof pen and do not use
after one year.
The color calibration set may no longer be used once either of the two
periods has expired.
The software notifies you that the color calibration set needs to be
replaced with a new set before the color calibration expires.
Once the color calibration set has expired the software notifies you that
a color analysis can only be carried out based on old calibration data.
Please contact your dealer for replacements for the color calibration set.
8 Maintenance
8.3 Color calibration