Installation - Swann AMU Powercon Operator's Manual

Version 6.20 and higher
Table of Contents

Advertisement

AMU Powercon
Program List and Explanations
4.3.1
Interval
1 s
Time
25 min
4.3.2

5 Installation

5.1 Sensors
5.1.1
5.1.2
5.1.2.1
5.1.2.2
5.1.2.3
5.1.2.4
A-96.250.331 / 230320
Log Interval: Select a convenient log interval. Consult the table be-
low to estimate the max logging time. When the logging buffer is
full, the oldest data record is erased to make room for the newest
one (circular buffer).
5 s
1 min
2 h
25 h
Clear Logger: If confirmed with yes, the complete logger data is de-
leted. A new data series is started.
Flow:
 None
 Q-Flow
Select "Q-Flow" if the sample flow should be monitored and shown
on the display and when using a SWAN flow cell.
Sensor parameters
Cell Constant: Enter the cell constant printed on the sensor label.
Range: 0.005000 cm
Temp. Corr: Enter the temperature correction printed on the sensor
label.
Range: -2 °C to 2 °C
Cable length: Enter the cable length.
Range: 0.0 m to 30.0 m
Meas. unit: Select the measuring unit as S/cm or as S/m.
5 min
10 min
5 d
10 d
-1
-1
–11.00 cm
30 min
1 h
31 d
62 d