Swann AMU Powercon Operator's Manual page 43

Version 6.20 and higher
Table of Contents

Advertisement

AMU Powercon
Error List
Error
E001
Cond. Alarm high
E002
Cond. Alarm low
E007
Sample Temp. high
E008
Sample Temp. low
E009
Sample Flow high
E010
Sample Flow low
E011
Temp. shorted
E012
Temp. disconnected
E013
Case Temp. high
E014
Case Temp. low
E017
Control time-out
E018
Quality Assurance
A-96.250.331 / 230320
Description
Corrective action
– check process
– check programmed value, see
p. 53
– check process
– check programmed value, see
p. 53
– check process
– check programmed value, see
p. 53
– check process
– check programmed value, see
p. 53
– check sample inlet pressure
– check programmed value, see
5.3.1.2.2, p. 53
– check sample inlet pressure
– Check flow regulating valve
– check programmed value, see
5.3.1.2.35, p. 53
– Check wiring of temperature sensor
– Check temperature sensor
– Check wiring of temperature sensor
– Check temperature sensor
– check case/environment temperature
– check programmed value, see
p. 54
– check case/environment temperature
– check programmed value, see
p. 54
– Check control device or programming
in Installation, Relay contact, Relay 1/2
5.3.2/3, p. 54
– Perform QA Procedure using reference
instrument, e.g. AMI Inspector
5.3.1.1,
5.3.1.1,
5.3.1.3,
5.3.1.3,
5.3.1.4,
5.3.1.5,
37

Advertisement

Table of Contents
loading

Table of Contents