AMU Powercon
Program List and Explanations
5.2.1.3
As process
values
5.2.1.40
A-96.250.331 / 230320
Function: Define if the signal output is used to transmit a process
value or to drive a control unit. Available functions are:
Linear, bilinear or logarithmic for process values.
See
As process values, p. 49
Control upwards or control downwards for controllers.
See
As control output, p. 50
The process value can be represented in 3 ways: linear, bilinear or
logarithmic. See graphs below.
[mA]
20
B
10
12
0 / 4
0.0
0.1
0.2
A
linear
B
bilinear
[mA]
20
10
12
0 / 4
2
4
6
1
10
100
0
1
2
X Measured value (logarithmic)
Scaling: Enter beginning and end point (Range low & high) of the
linear or logarithmic scale. In addition, the midpoint for the bilinear
scale.
A
0.3
0.4
0.5
X Measured value
X
1'000
10'000
3
4
X
49