Swann AMU Powercon Operator's Manual page 58

Version 6.20 and higher
Table of Contents

Advertisement

AMU Powercon
Program List and Explanations
5.2.1.43
5.2.1.43.12
5.2.1.43.22
5.2.1.43
5.2.1.43.13
5.2.1.43.23
5.2.1.43.3
5.2.1.43.4
5.2.1.43.5
5.3 Relay Contacts
5.3.1
52
Control Parameters: if Parameters = Sample flow
Setpoint
Range: 0 –50 l/h
P-Band:
Range: 0 –50 l/h
Control Parameters: if Parameters = Cond. uc.
Setpoint
Range: 0 S–300 mS
P-Band:
Range: 0 S–300 mS
Reset time: The reset time is the time till the step response of a sin-
gle I-controller will reach the same value as it will be suddenly
reached by a P-controller.
Range: 0–9'000 sec
Derivative time: The derivative time is the time till the ramp re-
sponse of a single P-controller will reach the same value as it will
be suddenly reached by a D-controller.
Range: 0–9'000 sec
Control timeout: If a controller action (dosing intensity) is constantly
over 90% during a defined period of time and the process value
does not come closer to the setpoint, the dosing process will be
stopped for safety reasons.
Range: 0–720 min
Alarm Relay: The alarm relay is used as cumulative error indicator.
Under normal operating conditions the contact is active.
The contact is inactive at:
 Power loss
 Detection of system faults like defective sensors or electronic
parts
 High case temperature
 Process values out of programmed ranges.
Program alarm levels, hysteresis values and delay times for the fol-
lowing parameters:
 Alarm Conductivity
 Sample Flow
 Sample Temp.
 Case Temp. high
 Case Temp. low
A-96.250.331 / 230320