Product Description; Description Of The System - Swann AMU Powercon Operator's Manual

Version 6.20 and higher
Table of Contents

Advertisement

AMU Powercon

Product Description

1.
1.1.
Application
Range
Special
Features
Signal
Outputs
Relays
A-96.250.331 / 230320
Product Description

Description of the System

This instrument is applicable for the measurement of conductivity in
high purity water applications.
The AMU Powercon measures the specific (total) conductivity or
the acid (cation) conductivity of a sample. For the measurement of
the acid conductivity a cation exchanger bottle is necessary. The
transmitter can be used with a two-electrode conductivity sensor
with an integrated Pt1000 temperature sensor, e.g. Swansensor
UP-Con1000.
The conductivity is a parameter for the total quantity of ions present
in the solution. It can be used for the controlling of:
 the condition of waters
 water purification
 water hardness
 completeness of ion analysis
Many temperature compensation curves for specific conductivity
measurement:
 none
 Coefficient
 Neutral salts
 High-purity water
 Strong acids
 Strong bases
 Ammonia, Eth. am.
 Morpholine
Two signal outputs programmable for measured values (freely scal-
able, linear, bilinear, log) or as continuous control output (control
parameters programmable).
Current loop:
Maximal burden:
Two potential-free contacts programmable as limit switches for
measuring values, controllers or timer for system cleaning with au-
tomatic hold function.
Maximum load: 100 mA/50 V
0/4–20 mA
510 Ω
3

Advertisement

Table of Contents
loading