The Flow cell Catcon+ SL is used for the measurement of the acid
(cation) conductivity. Only one sensor can be used together with
the AMU Powercon. Therefore the left hand bore has to be closed
with a blind plug.
5.3
38
235
 Temperature: up to 50 °C
 Inlet pressure: 0.2 to 2 bar
 Outlet pressure: ambient pressure
 Sample flow: 5 to 20 l/h
 Inlet: ¼" Swagelok tube adapter (SS)
 Outlet: 8/6 mm Serto tube adapter (PVDF)
The maximal recommended cable length is 15 m, select the instal-
lation location accordingly. The flow cell must be aligned in exactly
vertical position.