AMU Powercon
Installation
Catcon+ SL
Sample
conditions
Sample
connection
Installation
A-96.250.331 / 230320
Example 2
The Flow cell Catcon+ SL is used for the measurement of the acid
(cation) conductivity. Only one sensor can be used together with
the AMU Powercon. Therefore the left hand bore has to be closed
with a blind plug.
5.3
38
235
Temperature: up to 50 °C
Inlet pressure: 0.2 to 2 bar
Outlet pressure: ambient pressure
Sample flow: 5 to 20 l/h
Inlet: ¼" Swagelok tube adapter (SS)
Outlet: 8/6 mm Serto tube adapter (PVDF)
The maximal recommended cable length is 15 m, select the instal-
lation location accordingly. The flow cell must be aligned in exactly
vertical position.
62
361
13