AMU Powercon
Program List and Explanations
5.3.4.5
5.4 Miscellaneous
5.4.1
5.4.2
5.4.3
5.4.4
5.4.5
5.5 Interface
5.5.1
5.5.20
5.5.30
5.5.40
5.5.1
5.5.21
5.5.31
5.5.41
5.5.1
58
Delay: Time which the instrument waits, after the input is deactivat-
ed, before returning to normal operation.
Range: 0–6'000 Sec
Language: Set the desired language.
Available settings: German /English/French/Spanish/Italian
Set defaults: Reset the instrument to factory default values in three
different ways:
 Calibration: Sets calibration values back to default. All other
values are kept in memory.
 In parts: Communication parameters are kept in memory. All
other values are set back to default values.
 Completely: Sets back all values including communication
parameters.
Load Firmware: Firmware updates should be done by instructed
service personnel only.
Password: Select a password different from 0000 to prevent unau-
thorized access to the menus "Messages", "Maintenance", "Opera-
tion" and "Installation".
Each menu may be protected by a different password.
If you forgot the passwords, contact the closest SWAN representa-
tive.
Sample ID: Identify the process value with any meaning full text,
such as KKS number.
Select one of the following communication protocols. Depending on
your selection, different parameters must be defined.
Protocol: Profibus
Device address:
ID-Nr.:
Local operation:
Protocol: Modbus RTU
Device address:
Baud Rate:
Parity:
Protocol: HyperTerminal
Baud Rate:
Range: 0–126
Range: Analyzer; Manufacturer; Multivariable
Range: Enabled, Disabled
Range: 0–126
Range: 1 200–115 200 Baud
Range: none, even, odd
Range: 1200–115200 Baud