Output/Control; Input Active; Signal Outputs - Swann AMU Powercon Operator's Manual

Version 6.20 and higher
Table of Contents

Advertisement

AMU Powercon
Program List and Explanations
5.3.2.6
5.3.2.7
5.3.2.1
5.3.4
5.3.4.1
5.3.4.2
5.3.4.3
5.3.4.4
A-96.250.331 / 230320
Signal Outputs: select the behavior of the signal outputs when the
relay closes. Available values: cont., hold, off
Output/Control: select the behavior of the control outputs when the
relay closes. Available values: cont., hold, off
Function = Fieldbus:
The relay will be switched via the Profibus input. No further param-
eters are needed.
Input: The functions of the relays and signal outputs can be de-
fined depending on the position of the input contact, i.e. no function,
closed or open.
Active: Define when the input should be active:
No:
Input is never active.
When closed
Input is active if the input relay is closed
When open:
Input is active if the input relay is open
Signal Outputs: Select the operation mode of the signal outputs
when the relay is active:
Continuous:
Signal outputs continue to issue the measured
value.
Hold:
Signal outputs issue the last valid measured value.
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Off:
Set to 0 or 4 mA respectively. Errors, except fatal
errors, are not issued.
Output/Control: (relay or signal output):
Continuous:Controller continues normally.
Hold: Controller continues on the last valid value.
Off:
Controller is switched off.
Fault:
No:
No message is issued in pending error list and the
alarm relay does not close when input is active.
Message E024 is stored in the message list.
Yes:
Message E024 is issued and stored in the mes-
sage list. The Alarm relay closes when input is
active.
57

Advertisement

Table of Contents
loading

Table of Contents