Swann AMU Powercon Operator's Manual page 60

Version 6.20 and higher
Table of Contents

Advertisement

AMU Powercon
Program List and Explanations
5.3.1.3.25
5.3.1.4
5.3.1.5
5.3.2/3
5.3.2.1
5.3.2.20
5.3.2.300
54
Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E008 is displayed in the message
list.
Range: -10 to +20 °C
Case Temp. high: Set the alarm high value for temperature of elec-
tronics housing. If the value rises above the programmed value
E013 is issued.
Range: 30–75 °C
Case Temp. low: Set the alarm low value for temperature of elec-
tronics housing. If the value falls below the programmed value
E014 is issued.
Range: -10 to +20 °C
Relay 1 and 2: The function of relay contacts 1 or 2 is defined by
the user.
NOTICE: The navigation in the menu <Relay 1> and <Relay 2>
is equal. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1 First select the functions as:
- Limit upper/lower,
- Control upwards/downwards,
- Timer
- Fieldbus
2 Then enter the necessary data depending on the selected func-
tion. The same values may also be entered in menu
Contacts, p. 46
Function = Limit upper/lower:
When the relays are used as upper or lower limit switches, program
the following:
Parameter: select a process value
Setpoint: If the measured value rises above respectively falls below
the set-point, the relay is activated.
Parameter
Range
Conductivity
0 S–300 mS
Temperature
-25 to + 270 °C
Sample flow
0–50 l/h
Cond. uc
0 S–300 mS
4.2 Relay
A-96.250.331 / 230320