Compressor; Remove And Replace The Compressor Box - CAIRE 1000B Provider Technical Maunal

Eclipse3 with autosat personal ambulatory oxygen system
Table of Contents

Advertisement

Compressor

As with any concentrator, the compressor is a limited-wear component and may require servicing
during the lifetime of the device. The point of service will be dependent on factors such as operation
time, flow settings and environmental conditions. Service is required when O
cannot be maintained.

REMOVE AND REPLACE THE COMPRESSOR BOX

NOTE:
1.
Remove the Unit Cover as described in the section Remove and Replace the Unit Cover.
CAUTION
2.
Flip unit over onto a padded ESD safe surface. Using a small flat tip screwdriver to pry open the cover at each
pry point, remove the Case Bottom Cover as shown in Figure 66.
Personal Ambulatory Oxygen System Provider Technical Manual
There are no field serviceable parts inside the Compressor Box. Do not
attempt to disassemble or modify the Compressor Box in the field.
The Eclipse contains electrostatic sensitive components. Do not open or
handle except at a static free workstation.
Case Bottom
Cover
Figure 71: Case Bottom Cover Removal
purity and/or flow rates
2
Pry
Point
Pry
Point
Pry
Point