Eclipse Oxygen System Specifications - CAIRE 1000B Provider Technical Maunal

Eclipse3 with autosat personal ambulatory oxygen system
Table of Contents

Advertisement

ECLIPSE OXYGEN SYSTEM SPECIFICATIONS

Oxygen Concentrator
Dimensions (H x W X D)
Weight
Eclipse
Power Cartridge
Flow Settings
Continuous Flow
(measured in Liters Per Minute LPM)
Pulse Dose (measured in mL)
Continuous Flow Accuracy
Oxygen Concentration
Maximum System Pressure
Oxygen Output Pressure
Oxygen Concentration
Status Indicator
Nominal Sound Level
3.0 LPM Continuous Flow
3.0 Pulse Dose Mode
Operating Environment
Temperature
Humidity
Storage Environment
Temperature
Humidity
Altitude
Nominal Power
0.5 LPM Continuous flow
3 LPM Continuous Flow
1.0 Pulse Dose Setting
6.0 Pulse Dose Setting
Battery Charging
Nominal Power Cartridge operating time
Continuous Flow
2.0 LPM
3.0 LPM
Pulse Dose (@ 12 BPM)
2.0
6.0
Continuous Flow Indication
Audible Alarm Indicators
Back-Up Alarm Power
Filters
Device Classification
Personal Ambulatory Oxygen System Provider Technical Manual
19.3 x 12.3 x 7.1 inches
(49.0cm x 31.2cm x 18.0cm)
15.0 pounds
3.4 pounds
0.5 to 3.0 LPM (0.5 liter increments)
16-96 (16mL increments), 128mL, 160mL, 192mL
+/- 10% or 200ml/min, whichever is greater
87–95.6% for all flow settings
15 psig (103.5 kPa)
5.0 psig (34.5 kPa) nominal
Green Light = Normal Operation
Yellow Light = Warning or Caution, less than 85% ± 3%
Red Light Flashing = Abnormal Operation, less than 70% ± 5%
48 dB(A)
40 dB(A)
50º F to 104º F (+10° to 40°C)
10% to 95%, Non-condensing, 82.4°F (28°C) Maximum Dew point
-4º F to 140º F (-20° to 60°C)
Up to 95% Non-condensing
0 – 13,123 feet (0 - 4,000 meters)
48 Watts
145 Watts
45 Watts
95 Watts
50 Watts
2.0 hours
1.3 hours
Note: Times will decrease with higher bolus size,
breath rate, ambient temperature, power cartridge
5.2 hours
age and use over time.
3.5 hours
Expressed in liters per minute (LPM)
• Loss of Power/Hot Power Cartridge
• Low Power Cartridge/Warm Power Cartridge
• Low Oxygen Output
• O
Flow Outside Normal Limits
2
• Unit Malfunction
• Low 9V battery
9V Internal Battery
Air Inlet, HEPA, Compressor Intake
IEC Class I, Type B Applied Part, IPX0
6