12. Remove caps from ATF Module.
13. Connect the silicone tube from the Product Tank into the product port of the ATF and secure joints with
cable ties as shown on Figure 30 on page 57.
14. Use cable tie to secure the silicone tube to the braid tube to ensure that there is NO kinking on the
bend as shown on Figure 31 on page 57. Hand-tighten the cable tie. Inspect the tubing to ensure the
tubing is not kinked.
15. Perform product test prior to installing cover and again after cover replacement.
16. Installation of the Unit Cover on the unit is described in the Remove and Replace of the Unit Cover.
REMOVE AND REPLACE THE PRODUCT TANK ASSEMBLY (PN
6138-SEQ)
ECLIPSE 1, ECLIPSE 2, EARLY ECLIPSE 3
1.
Remove the Unit Cover as described in the section Remove and Replace the Unit Cover.
2.
Cut the 2 green cable ties that secure the silicone tubes to the top of the Product Tank as shown in
Figure 35 below. Disconnect these two silicone tubes. Pinch the silicone tube coming from the ATF
product port and secure with a cable tie as shown on Figure 36 on next page.
from ATF Module
Personal Ambulatory Oxygen System Provider Technical Manual
Silicone Tube from
Flow/Concentration
Measurement Tube
Silicone Tube
Product Port
Figure 40: Product Tank inlet and Outlet Connections.
Cable
Tie