CAIRE 1000B Provider Technical Maunal page 59

Eclipse3 with autosat personal ambulatory oxygen system
Table of Contents

Advertisement

Battery Power: My
Battery is not Lasting
the time it used to...
NOTE: Battery times
will decrease with
higher bolus size,
breath rate, ambient
temperature, Battery
age and use over time.
Eclipse runs on Battery
power when plugged
into external power
(AC or DC)
Unit will beep for loss
of external power
Power Cartridge
(Battery) status gauge
no longer indicates
fully charged
WARNING:
Personal Ambulatory Oxygen System Provider Technical Manual
OTHER CONDITIONS (CONTINUED)
Battery End of Life: May have exceeded
the life expectancy of 500 full Charge/
Discharge cycles.
Calibration: The Eclipse Battery may be
out of calibration
The Eclipse display may display (****)
stars in place of the battery display
A power interruption that causes The
Eclipse display to read "FAIL 81", or
"FAIL 95"
*Red Light and constant tone alert
Other
AC power is loose or not connected
DC power is loose or not connected
Unit runs the battery down when you
don't expect it to?
Other
Power Cartridge (battery) is partially/
completely discharged.
Power Cartridge (battery) is not seated
properly in the Eclipse compartment.
Power Cartridge (battery) out of
Calibration.
Power Cartridge (battery) is aging.
Other
WARNING: Do not modify this equipment without authorization of the
manufacturer.
Discharge it fully, recharge it fully, then measure the length of discharge to
within 20% of new specification. Replace if higher that 20% loss of charge.
*Unplug the Eclipse from external
*Allow Eclipse to operate until the Red Light comes on and alarm sounds
constant(Loss or Power)
*Plug in to External AC power and allow to recharge
Indicates a loss of communication from the battery. Change the Battery
with a fresh one. If the issue remains, contact your home oxygen provider
for service.
This error code may indicate a Power Loss and reconnection. *Remove all
power(Ext power and Battery cartridge) from the Eclipse. *Wait 15 Seconds.
*Re-Connect External AC power and listen for the External Power connec-
tion "3 second beep". Press the "ON" button and allow the Eclipse to come
up to Green Light operation.
Contact Chart TS for RMA service
Plug device directly into known good AC outlet, extension cord and com-
puter strips are not recommended
*Remove battery from the eclipse and ensure can operate the Eclipse un-
interrupted on car power
*Use power cartridge(s) if car does not have adequate power
*Have outlet checked for proper current rating
*Unit may be plugged into external power but the power verification light
is "not" on
*Check outlet, check plug on power supply, check power plug on Eclipse
Contact Chart TS for RMA service
Plug device directly into known good AC outlet (verify a waterfalling
[scrolling bottom to top] battery indicator)
*Reseat Power Cartridge in the Eclipse battery compartment
*Ensure both locking pins are engaged [check for waterfalling (scrolling
bottom to top) battery indicator if not fully charged]
Calibrate Battery: Unplug ext power turn unit on, allow to run until
power loss (Red Light and constant tone alert), plug unit into AC power to
charge (ensure indicator is waterfalling [Scrolling Bottom to top]).
Typical average end of life is 500 charge/discharge cycles: Replace
with a known good battery
Contact Chart TS for RMA service.