CAIRE 1000B Provider Technical Maunal page 58

Eclipse3 with autosat personal ambulatory oxygen system
Table of Contents

Advertisement

Symptom
During Operation:
Red Light "Blink-
ing" and "3 Beeps"
(Low O2 condition
- Less than 70%)
Press NO SMOKING
button for Diagnostic
Code "ALRM=004"
During Operation:
Red Light
"Constant" and alarms
"constant tone"
FAIL Condition
Displayed "FAIL XX"
Eclipse Red Light
and Alarms: Battery
has become Too Hot.
FAIL Condition
Displayed "FAIL XX"
Symptom
Pulse Mode: Green
Pulse Mode Led is
Steady then Flashing
Press NO SMOKING
button for Diagnos-
tic Code "ALRM=200"
(Eclipse is going from
Pulse to Continuous
every 15 seconds.)
Personal Ambulatory Oxygen System Provider Technical Manual
Probable Cause
Restricted Air Flow through cabinet
inlet filter
Blocked air flow to cabinet inlet
filter
Restricted air flow through Internal
Filters
Possible Blockage or Leak of air
flow through internal components
Unit has shut down and displays a
FAIL XX code
Other
Battery is too Hot and will not
operate the Eclipse; above 158°F
(70°C)
Eclipse Overheated, Display shows
a FAIL 92, Fail 91, or FAIL A4
Other
Probable Cause
Not detecting a patient's breath
Unit is in a Breath Search mode
Other
RED LIGHT
Perform Annual Preventive Maintenance (5022-SEQ), if issue remains call
Chart TS for service.
Ensure the Eclipse has adequate air flow around the device (6" around the
device) - turn the device around in the car(exhaust facing to the middle of
the car)
Perform Annual Preventive Maintenance (5022-SEQ), if issue remains call
Chart TS for service.
Contact Chart TS for RMA service.
May indicate a Power Loss Condition or Internal malfunction.
*Remove all power (Ext power and Battery cartridge) from the Eclipse.
*Wait 15 Seconds.
*Re-Connect External AC power and listen for the External Power connec-
tion "3 second beep". Press the "ON" button and allow the Eclipse to come
up to Green Light operation.
Contact Chart TS for RMA service
*Remove the battery and allow to cool
*Replace with a fresh cool battery.
Allow the Eclipse to cool off (30 Minutes Minimum) below the internal
temperature of 115 F - remove power, wait 15 seconds, plug back in to
External Power only and restart the Eclipse. Install battery if needed.
Contact Chart TS for RMA service.
OTHER CONDITIONS
Ensure Cannula Tubing is not more than 7ft (2.1m) in length. Ensure it is not
kinked or disconnected.
Ensure you are not using a Humidifier - check for water block in cannula
Ensure Patient is breathing through their nostrils. Patient may be a mouth
breather, not allowing enough negative pressure to trigger the device's
bolus volume.
Provider to properly set the Pulse Sensitivity settings for their client. (menu:
PS=XX)
Contact Chart TS for RMA service
Remedy
Remedy
58